Legionella Checks in Dental Practices: A Practical CQC Compliance Guide

Writer: Liam
Liam
Oct 22
13 min read
Dental nurse flushing dental unit waterlines to reduce Legionella risks.
Effective dental waterline management includes regular flushing, disinfection and maintenance in accordance with the practice’s risk assessment and manufacturer guidance.

Every dental practice must assess and manage the risk of Legionella bacteria within its water systems. This includes the building’s hot and cold water supplies and the dental unit waterlines used during patient treatment.

Dental unit waterlines, commonly abbreviated to DUWLs, require particular attention because they consist of narrow plastic tubing through which water may flow slowly or remain stagnant. These conditions can encourage bacteria and biofilm to develop.

Dental treatment can also generate aerosols. If waterlines are not properly maintained, contaminated water droplets may be released into the surgery, potentially exposing patients and members of the dental team.

Good Legionella management is therefore not simply about obtaining a risk assessment. A practice must have a suitable written waterline management scheme, follow the dental chair manufacturer’s instructions, complete the required checks and act on recommendations.


What is Legionella?

Legionella is a type of bacterium found in natural and purpose-built water systems. If contaminated water is released as fine droplets or aerosols, people may inhale the bacteria.

Legionella can cause Legionnaires’ disease, a potentially fatal form of pneumonia. Although anyone can become infected, some people are at greater risk, including:

  • People over 45

  • Smokers and heavy drinkers

  • People with chronic respiratory or kidney disease

  • People with diabetes, lung disease or heart disease

  • People with a weakened immune system

Dental practices treat people from all of these groups. Some patients may also be particularly susceptible because of age, illness or medical treatment.

The HSE Approved Code of Practice L8 explains that Legionella growth is more likely where:

  • Water temperatures remain between 20°C and 45°C

  • Water is stored or recirculated

  • Water outlets are used infrequently

  • Dead legs or redundant pipework allow stagnation

  • Scale, rust, sludge or biofilm is present

  • Equipment produces aerosols or water droplets

In dental practices, the small diameter and construction of dental unit waterlines introduce additional risks.


Why are dental unit waterlines a particular concern?

Dental unit waterlines supply water to equipment such as:

  • Dental handpieces

  • Ultrasonic scalers

  • Three-in-one syringes

  • Cup fillers

  • Other water-cooled dental equipment

The narrow plastic tubing found inside a dental unit provides conditions in which biofilm can form. Biofilm is a layer of microorganisms attached to the inside of a surface, protected by a slimy matrix.

CQC identifies several reasons why biofilm can develop quickly in dental waterlines:

  • The water can contain nutrients for bacteria

  • Plastic tubing may contain substances that support bacterial growth

  • Surgery temperatures can warm the water

  • Water may remain stagnant overnight, during lunch breaks and at weekends

  • Water flow through narrow tubing may be relatively slow

Once established, biofilm can be difficult to remove. Flushing helps reduce the number of microorganisms present in the water at the time of use, but flushing alone will not reliably eliminate established biofilm.

This is why dental practices normally need a combination of measures, including flushing, suitable disinfection, correct water management and regular maintenance.


What does CQC expect from dental practices?

CQC Dental Mythbuster 5 sets out several mandatory requirements for dental practices.

According to CQC:

  1. All relevant water systems require a risk assessment, although some may only need simple control measures.

  2. Dental premises must have a written waterline management scheme and Legionella risk assessment.

  3. These documents must be prepared by experienced and competent people.

  4. The registered manager must implement the recommendations from the risk assessment and waterline management scheme.

  5. Water and air lines must have suitable anti-retraction valves.

  6. The practice must control Legionella within its dental waterline system, although no single treatment will be completely effective.

CQC considers Legionella and dental waterline management when assessing whether the practice provides a safe environment. Its dental guidance specifically relates this to Regulation 15: Premises and equipment.

Failures may also have wider implications under Regulation 12, Safe care and treatment, and Regulation 17, Good governance, particularly if risks are not acted upon or monitoring records are incomplete.

The practice therefore needs to demonstrate more than possession of a contractor’s report. It should be able to show that:

  • Recommendations have been completed

  • Daily waterline procedures are followed

  • Responsibilities are clearly allocated

  • Dental chairs are maintained correctly

  • Staff understand the flushing and disinfection procedure

  • Problems are identified and escalated

  • Records provide reliable evidence of compliance


Which guidance applies to dental practices?

Dental practices should consider several pieces of national guidance:

HTM 01-05 contains specific recommendations for dental unit waterlines. However, the entire building water system must also be considered. A practice should not focus on the dental chairs while overlooking stored water, poorly used taps or inappropriate pipework elsewhere in the building.


Who is responsible for Legionella control?

The legal dutyholder is normally the employer or the person who controls the premises or water system.

Within a dental practice, the registered manager should ensure that suitable arrangements are established and maintained. A responsible person should be appointed to oversee Legionella controls.

The practice should document who is responsible for:

  • Arranging the Legionella risk assessment

  • Maintaining the water system

  • Managing dental unit waterlines

  • Flushing waterlines

  • Disinfecting dental waterlines

  • Cleaning independent water bottles

  • Monitoring water temperatures

  • Checking low-use outlets

  • Arranging maintenance and repairs

  • Reviewing contractor reports

  • Recording and closing remedial actions

  • Reviewing the waterline management scheme

Where the premises are leased, the landlord may be responsible for parts of the building water system. The practice may remain responsible for the dental chairs, waterline procedures and equipment used to provide care.

These responsibilities should be confirmed in writing. The practice should obtain evidence that landlord-controlled checks and maintenance have been completed rather than simply assuming that they have.

Using a contractor does not remove the dutyholder’s responsibility. The practice must take reasonable steps to ensure that the contractor is competent and that recommendations have been acted upon.


Does an external specialist need to complete the assessment?

CQC and HSE require the person completing the Legionella risk assessment to be competent. Competence means having the necessary knowledge, skills, training and experience to assess the relevant system.

HTM 01-05 recognises that dental waterline management is a specialised field and advises registered managers to obtain support from an external specialist where necessary.

Specialist advice is particularly important where:

  • The building has stored water

  • The plumbing system is large or complex

  • Several organisations share the premises

  • Water temperatures repeatedly fall outside required parameters

  • The practice has old or poorly documented pipework

  • Dead legs or redundant pipes may be present

  • There is visible contamination within water bottles or tubing

  • A waterline treatment regime is not working

  • Legionella or high bacterial counts have been detected

  • The practice is installing or replacing dental chairs

  • The practice is unclear about backflow protection

  • The premises have been closed for a prolonged period

The assessment should cover both the building water system and the dental equipment connected to it.


What should a dental Legionella risk assessment include?

A suitable assessment should consider:

  • The source of the incoming water

  • Hot and cold water distribution

  • Water heaters, cylinders and storage tanks

  • Pipework design and condition

  • Sentinel outlets

  • Low-use taps, showers and other outlets

  • Dead legs and redundant pipework

  • Thermostatic mixing valves

  • Dental chairs and their water supplies

  • Independent water bottles

  • Dental unit waterlines

  • In-line or disposable filters

  • Anti-retraction valves

  • Ultrasonic scalers and handpieces

  • Three-in-one syringes

  • Spittoons and cup fillers

  • Wet-line suction systems

  • Equipment capable of producing aerosols

  • Patients who may be particularly susceptible

  • Waterline cleaning and disinfection products

  • Temporary closure and recommissioning procedures

  • Staff responsibilities and competence

The assessment should lead to a clear written control scheme. This should tell staff exactly what needs to be done, by whom, how often and what records must be kept.


How often should dental unit waterlines be flushed?

HTM 01-05 recommends that dental unit waterlines are flushed:

  • For at least two minutes at the beginning of the working day

  • For at least two minutes at the end of the working day

  • For at least two minutes after a significant period of non-use, such as a lunch break

  • For at least 20 to 30 seconds between patients

Care should be taken to minimise splashing and aerosol formation during flushing.

Each practice should confirm that this regime is compatible with the dental chair manufacturer’s current instructions. Some dental units have automated flushing or water-treatment programmes, and these should be operated exactly as specified.

Flushing is useful because it reduces the number of microorganisms present in the water delivery tube at the time patients may be exposed. However, it has only a limited effect on established biofilm.

A flushing regime must therefore be supported by effective cleaning, disinfection and maintenance.


How should dental waterlines be disinfected?

CQC advises practices to use suitable waterline treatment products and follow the dental chair manufacturer’s instructions.

Not every product is compatible with every dental unit. Using an unsuitable chemical could:

  • Damage tubing or seals

  • Damage dental handpieces

  • Affect filter performance

  • Interfere with bonding materials

  • Cause corrosion

  • Leave unsafe chemical residues

  • Invalidate equipment warranties

The practice should obtain written manufacturer guidance covering:

  • Approved waterline products

  • Correct product concentration

  • Continuous or intermittent dosing

  • Required contact time

  • Shock-treatment procedures

  • Flushing after treatment

  • Filter replacement

  • Independent bottle cleaning

  • Compatibility with handpieces and other equipment

  • Required personal protective equipment

Staff should not introduce a disinfectant merely because it is marketed for dental waterlines. Its suitability for the particular dental unit must be confirmed.

Where a chemical is used, the practice should also maintain the relevant Control of Substances Hazardous to Health assessment, safety information and staff instructions.


How should independent dental water bottles be managed?

Independent water bottles can help separate the dental chair from the building’s mains water supply, but they do not remove the risk of contamination.

HTM 01-05 recommends that self-contained dental water bottles are:

  • Removed at the end of the working day

  • Flushed with distilled or reverse-osmosis water

  • Left open to air-dry overnight

  • Stored inverted to reduce contamination

Where visual contamination is present, the bottle and system should be treated using a suitable disinfectant followed by thorough rinsing. The dental unit manufacturer’s instructions must be followed.

Self-contained water supplies used with dental care systems should normally use distilled or reverse-osmosis water as specified in HTM 01-05.

Using distilled or reverse-osmosis water does not, by itself, prevent biofilm. The bottle, tubing and waterline system must still be cleaned, disinfected and maintained.

If visual contamination is repeatedly found, the practice should not simply clean the bottle and continue as normal. The waterline management regime should be reviewed and competent advice obtained.


What are anti-retraction valves?

Anti-retraction valves help prevent contaminated fluids from being drawn back into the dental unit waterline.

Backflow or retraction may allow saliva, blood or other material from a patient’s mouth to enter equipment or waterlines. This creates a potential cross-infection risk.

CQC identifies anti-retraction valves as a mandatory waterline requirement. HTM 01-05 states that equipment requiring backflow protection should have suitable anti-retraction valves incorporated into the handpiece, scaler or waterline system.

Relevant equipment may include:

  • Dental handpieces

  • Ultrasonic scalers

  • Three-in-one syringes

  • Dental spittoons

  • Wet-line suction systems

  • Cup-filling systems

  • Other equipment connected to the water supply

Anti-retraction valves must be:

  • Correctly fitted

  • Checked in accordance with manufacturer instructions

  • Regularly maintained

  • Included in servicing arrangements

  • Replaced when defective

The practice should retain servicing or maintenance evidence showing that these components have been considered.


What water should be used during dental surgery?

For dental surgical procedures involving surgical flaps or access into body cavities where irrigation is required, HTM 01-05 recommends:

  • Sterile water, or

  • Sterile isotonic saline

This should be provided from a separate single-use source.

Water supplied through a conventional dental unit waterline should not be assumed to be sterile, even if the waterline is routinely treated and maintained.

The practice’s surgical procedure protocol should make this distinction clear.


What routine Legionella checks should a dental practice complete?

The exact monitoring schedule must be based on the practice’s risk assessment, equipment and manufacturer instructions. A typical programme may include:

Frequency

Typical dental water-safety task

Between patients

Flush DUWLs for at least 20 to 30 seconds

Start of day

Flush DUWLs for at least two minutes

After lunch or significant non-use

Flush DUWLs for at least two minutes

End of day

Flush DUWLs for at least two minutes

End of day

Drain and clean relevant dental water equipment as instructed

End of day

Remove, flush, air-dry and invert independent water bottles

Daily or as specified

Check waterline treatment product and dosing arrangements

As specified by manufacturer

Disinfect or shock-treat dental waterlines

At least twice weekly, or more frequently if assessed

Flush sporadically used building outlets

Monthly

Check hot and cold water temperatures at identified sentinel points

Quarterly or as required

Clean and descale showerheads and hoses, where present

Six-monthly

Check mixed-temperature outlet delivery temperatures where required

Annually

Inspect cold-water storage tanks, where present

Annually

Check representative taps on a rotational basis

According to manufacturer

Inspect and maintain anti-retraction valves, filters and water-treatment equipment

After closure

Follow the documented dental-unit recommissioning procedure

This table should not replace a site-specific control scheme. The practice’s procedures may need to be more frequent depending on the risk assessment.


What are sentinel outlets?

Sentinel outlets are selected taps used to monitor whether the building’s hot and cold water systems are achieving appropriate temperatures.

They are normally selected based on their position within the water system. These may include:

  • The first and last outlets on a circulating hot-water system

  • Outlets closest to and furthest from a water heater or storage tank

  • Outlets serving particularly susceptible patients

  • Outlets known to have poor temperature performance

  • Taps on long or infrequently used branches

HTM 01-05 indicates that monthly checks should usually confirm:

  • Cold water reaches below 20°C after running for up to two minutes

  • Hot water reaches at least 50°C after running for one minute

These measurements apply to appropriate non-mixed outlets. Where thermostatic mixing valves are installed, measurements may need to be taken at the inlets rather than from the blended outlet.

The competent risk assessor should identify the correct monitoring points and applicable temperature requirements. Staff should not alter thermostatic controls or expose patients to a scalding risk simply to obtain a particular reading.


Does a dental practice need routine water testing?

Routine microbiological monitoring is not automatically required for every dental practice.

CQC states that routine microbiological monitoring for total viable counts is generally unnecessary unless there are taste or odour problems. The frequency of Legionella testing should depend on the water system, its components and the findings of the risk assessment.

Testing may be appropriate where:

  • The risk assessment recommends it

  • Water temperatures are not being achieved

  • The effectiveness of the control regime is uncertain

  • There is visible contamination

  • There are recurring taste or odour problems

  • A waterline treatment system appears ineffective

  • The practice has a particularly susceptible patient population

  • Contamination has previously been detected

  • A case of Legionnaires’ disease may be associated with the premises

  • The practice uses a specialist purification or dosing system requiring monitoring

If microbiological testing is undertaken, HTM 01-05 states that approved methods and a UKAS-accredited laboratory should be used. It specifically states that dip slides are not acceptable for this purpose.

A low total viable count does not prove the absence of Legionella. Testing should therefore support, rather than replace, effective waterline management.


What should happen when a dental surgery closes temporarily?

Dental practices must have a procedure for decommissioning and recommissioning dental waterlines after a temporary closure.

The dental unit manufacturer’s instructions should be followed. In the absence of suitable manufacturer guidance, HTM 01-05 advises that temporarily unused dental unit waterlines should be flushed, drained and left disconnected where practicable.

Independent water bottles should be:

  • Removed

  • Flushed using distilled or reverse-osmosis water

  • Left open to air-dry

  • Stored inverted

When recommissioning, HTM 01-05 describes a process that includes flushing, suitable disinfection, further flushing and checking filters and anti-retraction valves. However, practices should use the manufacturer’s current instructions wherever these are available.

A documented recommissioning check should be completed before the dental chair is returned to clinical use.


What records should the dental practice maintain?

A dental practice should be able to produce reliable evidence of its water-safety arrangements.

A practical Legionella and dental waterline file should contain:

  • Current Legionella risk assessment

  • Written waterline management scheme

  • Building water-system schematic

  • Dental chair and waterline details

  • Named responsible person

  • Contractor competence information

  • Daily DUWL flushing records

  • Waterline disinfection records

  • Independent bottle cleaning records

  • Water-treatment product records

  • Sentinel outlet temperature records

  • Low-use outlet flushing records

  • Anti-retraction valve maintenance evidence

  • Filter cleaning and replacement records

  • Dental chair servicing reports

  • Water sampling results, where applicable

  • Action plans from the risk assessment

  • Evidence that remedial actions were completed

  • Closure and recommissioning records

  • Staff training and competency records

  • Review and audit records

Practices should avoid introducing forms that produce unnecessary duplication. A concise daily surgery checklist may record DUWL flushing, bottle management and treatment-product checks together.


How often should the Legionella risk assessment be reviewed?

HSE does not set a single expiry date that applies to every Legionella assessment. It should be reviewed regularly and whenever there is reason to believe that it may no longer be valid.

A documented annual management review is a sensible way to check whether the assessment and waterline management scheme remain current. This does not necessarily mean paying for a completely new external assessment every year.

An earlier review should be completed following:

  • Installation or replacement of a dental chair

  • Changes to waterline products

  • Plumbing alterations

  • Building refurbishment

  • Addition or removal of water outlets

  • Changes to independent bottle systems

  • Repeated temperature failures

  • Visible contamination

  • Unsatisfactory water-testing results

  • Significant changes in room use

  • Prolonged closure

  • A suspected case of Legionnaires’ disease

  • Changes in landlord or maintenance responsibilities

  • Updated manufacturer instructions

The review should be recorded even when no changes are required.


Common Legionella and dental waterline mistakes

Common weaknesses include:

  • Having a Legionella risk assessment that ignores dental chairs

  • Having no written dental waterline management scheme

  • Failing to implement recommendations

  • Relying on flushing as the only control measure

  • Using a waterline chemical not approved by the chair manufacturer

  • Assuming distilled or reverse-osmosis water prevents biofilm

  • Leaving independent water bottles filled overnight

  • Failing to flush after lunch or between patients

  • Not checking or maintaining anti-retraction valves

  • Using dental unit water for surgical irrigation

  • Taking temperatures from the wrong side of a mixing valve

  • Paying for routine testing without understanding why it is required

  • Using dip slides as evidence of Legionella control

  • Failing to document recommissioning after closure

  • Keeping records without reviewing failures or trends

  • Assuming the landlord manages all water-related risks


Dental practice Legionella compliance checklist

Your practice should be able to answer yes to the following:

  • Is there a current Legionella risk assessment?

  • Does it cover the building water system and dental unit waterlines?

  • Is there a written waterline management scheme?

  • Is a responsible person clearly identified?

  • Are landlord and practice responsibilities documented?

  • Are dental chair manufacturer instructions available?

  • Are DUWLs flushed at the required times?

  • Are independent water bottles managed correctly?

  • Are approved waterline products used?

  • Are disinfection and dosing arrangements documented?

  • Are anti-retraction valves fitted, checked and maintained?

  • Are low-use outlets identified and flushed?

  • Are sentinel outlet temperatures monitored as required?

  • Is sterile water or saline used for relevant surgical irrigation?

  • Are closure and recommissioning procedures documented?

  • Have all actions from the risk assessment been completed?

  • Are staff trained and competent in their assigned tasks?

  • Are abnormal results investigated and recorded?

  • Is the assessment reviewed following changes?


Final thoughts

Legionella control in a dental practice requires more than a generic premises risk assessment.

The practice must consider the complete building water system, but it must also manage the specific risks created by dental unit waterlines. This means following an effective combination of flushing, disinfection, water-bottle management, equipment maintenance and backflow protection.

The strongest evidence for CQC is not simply a Legionella certificate. It is a clear and consistent record showing that the practice understands its water systems, follows manufacturer guidance, completes the required controls and acts when a problem is identified.

If your dental practice needs help reviewing its compliance systems, preparing for CQC registration or completing a mock inspection, visit our CQC compliance support page.

If you want an independent check of your practice, a dental CQC mock inspection can help you spot gaps before CQC does.

This article provides general information and is not a substitute for a site-specific Legionella risk assessment, equipment manufacturer instructions or advice from a competent water-safety professional.